MX7B
The MX7B is an active vacuum gauge (also called an analog vacuum gauge or a digital vacuum gauge) featuring the well-known Televac 7B cold cathode ionization gauge. This gauge provides robust, contamination-resistant measurement from 10-3 to 10-8 Torr. The Penning ionization gauge is easy to disassemble, making gauge replacement and cleaning stress-free while also saving you time and money.
- Description
- Specifications
- Features & Benefits
- Modules
- Additional Details
Description
Our engineers designed this cold cathode ionization gauge with end users in mind, with a particular focus on heat treat and vacuum furnace end users who wanted full range vacuum measurement from a durable and cleanable digital vacuum gauge or analog vacuum gauge. The MX7B Penning ionization gauge is designed for use with our digital micron gauges, the MX4A convection active vacuum gauge or MX2A thermocouple active vacuum gauge, for a combined range of 10-8 Torr to 1000 Torr.
This system is also a drop-in replacement for the Televac® MP7ER cold cathode active vacuum gauge.
| Operating range | 1*10-8 Torr to 1*10-3 Torr |
| Display | OLED |
| Digital communication | RS-485, USB (micro) |
| Analog Output | Configurable 0 to 10 V DC |
| Programmable setpoints | 2 |
| Setpoint 1 | Open collector |
| Setpoint 2 | Relay |
| Supply voltage | 22 V DC to 26 V DC |
| Maximum performance | 8 W |
| Calibration medium | Dry air or nitrogen |
| Overpressure | 150 PSI |
| Resolution of the digital output | 3 significant digits with exponent |
| Analog output resolution | 16 bits |
| Operating temperature | 0 °C to 50 °C |
| Storage | -20 °C to 60 °C |
| Bake-out temperature | 120 °C |
| Reaction time | ≤150 ms |
| Display Readable Distance | 3 m (10 ft) |
| Accuracy | |
| 1*10-7 to 1*10-3 Torr | ±30% of reading |
| Analog Output | ±10 mV |
| PIN | Description |
| 1 | Analog Out (Ground) |
| 2 | RS-485 A (-) |
| 3 | SP2 NC |
| 4 | SP2 NO |
| 5 | SP1 Open Drain |
| 6 | Do not connect |
| 7 | Do not connect |
| 8 | Supply Voltage (+) |
| 9 | Analog Out (0 to 10) V |
| 10 | RS-485 B (+) |
| 11 | SP2 Com |
| 12 | SP1 Source |
| 13 | Sensor Voltage Control (Ground) |
| 14 | Sensor Voltage Control (+) |
| 15 | Supply Voltage (Ground) |

The MX7B has two set points. Set point 1 is an N-Channel 60 V MOSFET open collector. It has a maximum current rating of 1 A.
Set point 2 is a relay with a maximum switching voltage of 220 V DC (250 V AC) and a maximum switching current of 2 A.
| Brass |
| Stainless Steel |
| Tophet |
| Kovar |
| Ceramic |
| Glass |
| Fluoropolymer Elastomer |
| Nickel |
| Solder |
| Case | Aluminium |
| Weight with sensor | 1 kg (2 lbs) |
| Weight without sensor | 0.3 kg (0.6 lbs) |

| Certified to CAN/CSA C22.2 No. 61010-1-12 |
| CE certified to EN 61010-1, EN 61236-1, EN 55011 |
| RoHS compliant |
| IP40 |
| EAC certified |
| ISO 17025 accredited calibration optional |
| NIST-traceable calibration optional |
- Highly Contamination Resistant Cold Cathode Ionization Meter Design
- Compact design
- Color OLED display
- Configurable 0 to 10 V analog output
- Setpoint relay and open connection for process control
- RS-485 and USB communication standard
- EthernetIP communication available via the MX Active Gauge EIP gateway
- Easy-to-clean Penning ionization meter for extended sensor life
- ISO 17025 accredited and NIST traceable calibrations optional
- Fast response time for calibration and service
| MX7B Cold Cathode Active Vacuum Gauge | Part Number |
|---|---|
| KF25 SST | 2-8940-KF25 |