MX7M

The new Televac MX7M is an active vacuum gauge (also known as an analog vacuum gauge or a digital vacuum gauge) with the new Televac 7M cold cathode ionization gauge, which has a wide measuring range from 5*10-11 to 1*10-2 Torr (if you want to convert this value to another unit of measurement, check out our handy unit converter an). This is the only cleanable cold cathode on the market that measures in the range of 10-11 Torr. The new triple inverted magnetron design improves the overall accuracy of the vacuum gauge by improving linearity while significantly reducing the stray magnetic field.

 

Clear

Description

This vacuum gauge has all-metal seals, minimizing outgassing while maintaining a compact overall design. The MX7M cold cathode gauge also has a variety of features including a configurable 0 to 10 V DC analog output, RS-485 communications (EthernetIP optional), two programmable set points (relays and open collectors), and a bright OLED display with a 180° viewing angle.

All features of this cold cathode gauge are accessible through the display and touch controls or through the digital communications. Note that we also have an EthernetIP gateway for this active gauge which enables EthernetIP communication with a Rockwell Automation Allen-Bradley PLC. Combining this cold cathode gauge with a rough vacuum gauge such as the MX2A thermocouple active vacuum gauge or MX4A convection active vacuum gauge will provide full range measurement from 5*10 ®®-11 Torr to 1000 Torr (slightly above atmosphere).

Operating range 5*10-11 Torr to 1*10-2 Torr
Display OLED
Digital communication RS-485, USB (micro)
Analog Output Configurable 0 to 10 V DC
Programmable setpoints 2
Setpoint 1 Open collector
Setpoint 2 Relay
Supply voltage 22 V DC to 26 V DC
Maximum performance 8 W
Calibration medium Dry air or nitrogen
Overpressure 150 PSI
Resolution of the digital output 3 significant digits with exponent
Analog output resolution 16 bits
Operating temperature 0 °C to 50 °C
Storage -20 °C to 60 °C
Bake-out temperature 250 °C
Reaction time ≤50 ms
Display Readable Distance 3 m (10 ft)
Accuracy
1*10-9 to 1*10-3 Torr ±30% of reading
Analog Output ±10 mV

 

PIN Description
1 Analog Out (Ground)
2 RS-485 A (-)
3 SP2 NC
4 SP2 NO
5 SP1 Open Drain
6 Do not connect
7 Do not connect
8 Supply Voltage (+)
9 Analog Out (0 to 10) V
10 RS-485 B (+)
11 SP2 Com
12 SP1 Source
13 Sensor Voltage Control (Ground)
14 Sensor Voltage Control (+)
15 Supply Voltage (Ground)
Electrical connections diagram (pin diagram) for the Televac® MX7M Cold Cathode Active Vacuum Gauge.

The MX7M has two set points. Set point 1 is an N-Channel 60 V MOSFET open collector. It has a maximum current rating of 1 A.

Set point 2 is a relay with a maximum switching voltage of 220 V DC (250 V AC) and a maximum switching current of 2 A.

 

Stainless Steel
Alumina Ceramic
Molybdenum
Kovar
Ag-Cu Braze Alloy
Copper

 

Case Aluminium
Weight with sensor 1 kg (2 lbs)
Weight without sensor 0.3 kg (0.6 lbs)
Dimensional drawing of the Televac® MX7M Cold Cathode Active Vacuum Gauge.

Certified to UL 61010-1
Certified to CAN/CSA C22.2 No. 61010-1-12
CE certified to EN 61010-1, EN 61236-1, EN 55011
RoHS compliant
IP40
EAC certified
ISO 17025 accredited calibration optional
NIST-traceable calibration optional
  • Highly Contamination Resistant Cold Cathode Ionization Meter Design
  • All metal seals result in low outgassing
  • Compact design
  • Color OLED display
  • Configurable 0 to 10 V analog output
  • Setpoint relay and open connection for process control
  • RS-485 and USB communication standard
  • EthernetIP communication available via the MX Active Gauge EIP gateway
  • Easy-to-clean ionization meter for longer sensor life
  • ISO 17025 accredited and NIST traceable calibrations optional
  • Fast response time for calibration and service
MX7M Cold Cathode Active Vacuum Gauge Part Number
KF16 SST 2-8950-KF16
KF25 SST 2-8950-KF25
KF40 SST 2-8950-KF40
CF40 (2.75″ conflat) fixed SST 2-8950-CF40