Able to support 4 active gauges and up to 3 passive gauges, extending the measurement range to 2×10-12 mbar.
- Description
- Technical Data
- Features
- Applications
Description
The unit is fully software controlled. The unit can be remotely controlled via one of available interfaces.
Supply voltage | 100 – 240 VAC, 50 – 60 Hz (power consumption max 200 W) |
---|---|
Measurement chanels | up to 7 (active channels: 4; passive channels: up to 3 – number defined at the time of order) |
Supported active gauges | CTR90, TTR91, TTR211, PTR225, PKR251/360/361, PCR280, PTR90, ITR90, ITR100, Baratron, ANALOG IN, PG105, ATMION, User defined* |
Supported passive gauges | IE414, IE514, UHV-24, UHV-24p, MKS Series 274, NUDE-UHV-8A, NR-F-UHV, G8130, BARION basic II, User defined* |
Measurement range: Passive channel Active channel |
2×12-12 mbar 5×10-10 mbar |
Display range | 2×10-12 – 2×105mbar |
Setpoints relay outputs | 10 (4xNO/NC, 6xNO) |
Relay outputs with configurable activation signal | 4 (Emission, Degas) |
Digital inputs | 4 |
Analog outputs | 4 (0-10V) |
Communication interface | RS232/485, Ethernet, PROFINET (option – specify at order) |
Communication protocol | MODBUS-TCP |
User interface | 7″ TFT display with touchscreen |
Dimensions [mm] | 212.6 x 128.4 x 260.3 mm (WxHxD), 1/2 of 3HE 19″ |
Weight (approx.) | 3.8 kg |
Measurement range from atmospheric to 2×10-12 mbar
• Pressure plots
• Pressure trend graphs
• Measurement filtering (low, med, hi)
• Controls almost all commercially available Bayard-Alpert gauges
• Degassing of Bayard-Alpert with power and time limit
• Supports Ir/W/Thoria filaments
• Bayard-Alpert overpressure protection
• Selectable measurement units: mbar, Torr, Pascal, Psia
• Gas specific correction with one customized setting
• Unit over-temperature protection
• 10 individually programmable set-points with threshold and hysteresis functions
• User-defined channel names
• Possibility of active gauges self-identification
• Charts module – real time pressure charts with data export possiblity (option)
• Pararell measurement for all channels
• Flux monitor
• Black box version available (option) – without 7″ TFT display
• Fully software controlled (option)
• Autosave function for saving your parameters/unit preset and applying them automatically after restart
• Labview compatible (option)
- Any high and ultra high vacuum systems
- Measurement of pressure (from atmosphere to UHV range)
- Bakeable and non bakeable systems