MX2A

The MX2A Active Vacuum Gauge (also called a thermocouple vacuum gauge, analog or digital vacuum gauge, or Pirani gauge) utilizes an easily replaceable Televac 2A thermocouple vacuum gauge. This vacuum gauge has a variety of features, including a pressure measurement range from 1*10-4 Torr up to 1000 Torr, RS-485 communications, two programmable set points, and a selectable analog vacuum gauge output.

 

Clear

Description

The bright OLED digital vacuum gauge display makes it easy to read from a distance. Options on this Pirani gauge can be set from four touch controls on the top of the unit or through RS-485 communications (or EthernetIP if you add in an EthernetIP gateway). This vacuum gauge offers an extensive selection of fittings, simple use, ease of sensor replacement, and low cost, making it an excellent choice for a variety of vacuum applications.

Operating range 1*10-4 Torr to 1000 Torr
Communication RS-485
Analog Output 7 selectable 0 V DC to 10 V DC
Programmable setpoints 2
Setpoint 1 Open collector
Setpoint 2 Relay
Supply voltage 22 V DC to 26 V DC
Maximum performance 8 W
Calibration medium Dry air or nitrogen
Overpressure 150 psi
Resolution of the digital output 2 significant digits with exponent
Analog output resolution 16 bits
Operating temperature 0 °C to 50 °C
Bake-out temperature
Brass (PN: 2-8910-110) 100 °C (electronics removed)
Stainless Steel (PN: All Others) 200 °C (electronics removed)
Reaction time ≤2 sec
Display Readable Distance 3 m (10 ft)
Accuracy
0.1 mTorr to 1 mTorr 0.1 mTorr resolution
1 mTorr to 10 mTorr ±1 mTorr
10 mTorr to 1 Torr ±10% of reading
1 Torr to 100 Torr ±30 of reading (vertical mount)
100 Torr to 1000 Torr ±10% of measured value (vertical mounting)
Analog Output ±10 mV

 

Graph of the analog output behavior for the Televac® Thermocouple Active Vacuum Gauge.

PIN Description
1 Analog Out (Ground)
2 RS-485 A (-)
3 SP2 NC
4 SP2 NO
5 SP1 Open Drain
6 Floating
7 Floating
8 Supply Voltage (0 to 10) V
9 Analog Out (0 to 10) V
10 RS-485 B (+)
11 SP2 Com
12 SP1 Source
13 Floating
14 Floating
15 Supply Voltage (-)
Diagram of the electrical connections (pinout) for the Televac® MX2A Thermocouple Active Vacuum Gauge.

*0S1<CR> Read vacuum measurement
*0R1<CR> Read units of measurement
*0R2<CR> Read SP1L and SP1H

Assume address 0

Full list of commands is available in the instruction manual.

 

The MX2A has two set points. Set point 1 is an N-Channel 60 V MOSFET open collector. It has a maximum current rating of 1 A.

Set point 2 is a relay with a maximum switching voltage of 220 V DC (250 V AC) and a maximum switching current of 2 A.

 

MX2A PN: -110, -116, -125 MX2A PN: All Others
Nickel Nickel
Glass Glass
Gold Gold
Tophet Tophet
Cupron Cupron
Solder Stainless Steel

 

Case Aluminium
Electrical connections D-Sub 15-pin male
Weight with Sensor (typical) 345 g (0.76 lbs)
Weight without sensor 181 g (0.40 lbs)
Dimensional drawing of the Televac® MX2A Thermocouple Active Vacuum Gauge.

Certified to CAN/CSA C22.2 No. 61010-1-12
CE certified to EN 61010-1, EN 61236-1, EN 55011
RoHS compliant
IP40
EAC certified
ISO 17025 accredited calibration optional
NIST-traceable calibration optional
  • Cost-effective vacuum gauge
  • Very compact vacuum gauge
  • Easy-to-read digital OLED vacuum gauge color display
  • Convection meter with excellent resolution and repeatability (Pirani)
  • Superior instrument-to-device performance between vacuum gauges
  • ISO 17025 accredited and NIST traceable calibrations optional
  • Excellent customer support for vacuum gauges
MX2A Thermocouple Active Vacuum Gauge Part Number
KF16 SST 2-8910-130
KF25 SST 2-8910-131
KF40 SST 2-8910-132
CF16 (1.33″ conflat) fixed SST 2-8910-150
CF40 (2.75″ conflat) fixed SST 2-8910-152