MX200 EthernetIP
MX200 EthernetIP Vacuum Controller (also known as Vacuum Control Unit or Vacuum Pressure Regulator) can be connected to up to 10 Televac passive vacuum gauges, including cold cathode gauges, convection gauges, diaphragm gauges, and thermocouple vacuum gauges.
- Description
- Specifications
- Features & Benefits
- Modules
- Additional Details
Description
MX200 EthernetIP also features analog outputs from 0 to 10 V DC for each vacuum gauge connected to the vacuum controller, digital Ethernet IP/USB communication for easy integration with a PC or PLC, and up to 8 optional setpoint relays that serve as vacuum pressure regulators for process control in your vacuum system.
The user interface of the MX200 vacuum controller includes six touch controls and a high-contrast OLED display with a viewing angle of 180° that shows the readings of up to 8 connected vacuum gauges simultaneously along with the selected units of measurement and the status of the setpoint relays. This vacuum controller supports multiple gas types such as nitrogen/air, argon, hydrogen, helium, neon, krypton, and carbon dioxide. The MX200 vacuum controller is a modular unit that allows you to choose a vacuum sensing solution tailored to your needs.
Operating range | 1*10-11 Torr to 10,000 Torr |
Communication | USB/EthernetIP |
Analog Output | 0 to 10 V DC (1 per sensor) |
Analog output resolution | 16 bits |
Programmable setpoints | 4 per module (max. 8) |
Setpoint Type | Relay |
Supply voltage | 115/230 V AC (2/1 A), 50-60Hz |
Maximum performance | 230 W |
Calibration medium | Dry air or nitrogen |
Operating temperature | 0 °C to 50 °C |
Storage | -20 °C to 70 °C |
Display Readable Distance | 5 m (16 ft) |
Maximum Controlled Sensors | 10 |
Maximum Sensors Displayed | 8 |
Reaction time | ≤10 ms |
P1 | EthernetIP RJ45 port 1 |
P2 | EthernetIP RJ45 port 2 |
PS | USB communications |
MS | Module status LED |
USB | USB communications |
Setup Lock | Front panel lock switch |
BTLD | Bootload button |
Connectors | (2) RJ45 |
Ports | 10/100 Mbit, half/full duplex |
DHCP | Enable or disable |
Class 1 Connections | |
Connection Type | Point-to-point, multicast |
Target to Originator Instance ID | 100 |
Data Size | 40 |
Originator to Target Instance ID | 150 |
Data Size | 4 |
Configuration Instance | 3 |
Supported Connections | 4 |
Requested Packet Interval | 1 to 3200 ms |
Trigger Types | Cyclic, change of state |
Priorities | Low, high, schedule, urgent |
Class 3 Connections | |
Service | 0xE (get), 0x10 (set) |
Class | 0xA2 |
Instance | See ADI table |
Data Value Attribute | 5 |
Supported Connections | 6 |
Requested Packet Interval | 100 to 10000 ms |
Connection Type | Application |
Priority | Low |
The MX200 has up to eight relay set points. The relays are located on two optional relay modules with four set points each. Individual relay statuses are displayed on the front panel. The heavy-duty relays have an SPDT (1 Form C) contact form and are rated for 24 V DC. They have a maximum current rating of 8 A at 250 V AC or 5 A at 30 V DC.
Case | Aluminium |
Electrical connections | Power cord (IEC 60320 C-13) |
Weight (Base Unit) | 2.2 kg (5 lbs) |
Certified to UL 61010-1 |
CE certified to EN61326-1:2006 |
Certified to CAN/CSA C22.2 No. 61010-1-12 |
EAC certified |
ISO 17025 accredited calibration optional |
NIST-traceable calibration optional |
- Rack mountable with highly configurable design
- USB included in every device for easy connection to a PC
- High-contrast OLED display with 180° viewing angle
- Selectable units of measurement (Torr, mbar, Pascal)
- Bootloader for field-upgradable firmware
- Gas conversion functions for nitrogen/air, argon, hydrogen, helium, neon, krypton, and carbon dioxide included
- Analog outputs (0 to 10 V DC) for each connected passive vacuum gauge
- User-configurable analog output ranges
- Excellent customer support
- ISO 17025 accredited and NIST traceable calibrations optional
- Enable or disable DHCP with programmable IP address
- Status LEDs for network/module status/port data activity
- 10/100 Mbit, half- or full-duplex port operation
- Secure web access for network and device data and settings
MX200 | Part Number |
---|---|
Set point quad relay | 2-6200-411 |
2A thermocouple, 2 channel (for spares, specify MM or MX200) | 2-6200-486 |
4A convection, 2 channel (for spares, specify MM or MX200) | 2-6200-415 |
7B cold cathode, 1 channel (for spares, specify MM or MX200) | 2-6200-227 |
7E, 7F, 7FC, 7FCS cold cathode, 1 channel | 2-6200-285 |
1F piezo-diaphragm, 2 channel | 2-6200-244 |
1E piezo-diaphragm, 2 channel | 2-6200-220 |
CDG, 2 channel, specify range for each channel | 2-6200-452 |