LX218
The device is a leak detector for detecting and measuring leaks in test objects. The device is suitable for leak detection using the vacuum method and the sniffer method.
Category: Leak Detection
- Description
- Accessories
- Part Number
- Specifications
- Additional Details
Description
Integrated high quality components enable first-class performance for absolute measurement accuracy at outstanding speed. The simple navigation and robust construction make the LX218 leak detector the ideal choice for companies in a wide range of industries. The helium leak detector has a start-up time of less than three minutes and offers a fast leak test and reaction time. A further advantage: With its vacuum mode and sniffing option, it is perfectly suited for efficient leak localization and leak size determination.
Functions:
- When using the vacuum method the test object is evacuated and subjected to helium or forming gas from the outside. To do this it is necessary to establish a vacuum connection between the device and the test object.
- When using the sniffer method an overpressure is established in the test object using helium or a forming gas. The test object is then inspected on the outside using a sniffer line.
| LX218 Helium Leak Detector | |
|---|---|
| Part Number | Description |
| 551-300 | Sniffer line 3m |
| 551-301 | Sniffer line 5m |
| 551-302 | Sniffer line 10m |
| 551-010 | Remote Control RC1000C, wired, with 4m coiled cable |
| 551-015 | Remote Control RC1000WL, wireless, incl. wireless transmitter |
| LX218 Helium Leak Detector | |
|---|---|
| Part Number | Description |
| 8200-000 | LX218; Oil-Sealed pump; 230V ±10%; 50Hz |
| 8200-001 | LX218; Oil-Sealed pump; 115V ±10%; 60Hz |
| 8200-002 | LX218; Drycompressing pump; 230V ±10%; 50Hz |
| 8200-003 | LX218; Drycompressing pump; 115V ±10%; 60Hz |
| 8200-004 | LX218; Without pump; 100-230V ±10%; 50/60Hz |
| Type | LX218 Helium Leak Detector |
|---|---|
| Lowest detectable leak rate for He | ‹ 5 x 10E-13 Pa m³/s |
| Ion source | Two long-life iridium filaments, yttrium-oxide coated |
| Testing method | Vacuum and sniffer leak detection |
| Detectable gases | 4He, 3He, H2 |
| Internal He test leak | 10E-8 Pa m³/s |
| Leak rate display | 10E-13 – 0.1 Pa m³/s |
| Time to operation | 3 min. |
| Response time | 0.5 s |
| Suction capacity: He | 2.5 l/s |
| Connecting flange | DN 25 ISO-KF |
| Inlet pressure, max. | 25 mbar |
| Ambient temperature | +10°C to +35°C in operation |
| Input: analog | Compact gauge |
| Input: digital | Start/stop, zero, calibration |
| Interface | RS-232-C, RS-485 |
| Output: analog | Leak rate signal 2 x 0-10 V, lin/log |
| Output: digital | Ready to start, ready to measure, leak, error, partial flow valve |
| Relay: switch voltage | 2 x 230 V AC |
| Relay: switch current | 3 A |