The CC-10 Wide Range Active Vacuum Gauge (also called Analog VacuumGauge, Digital Vacuum Gauge, or Vacuum Pressure Gauge) uses two internal sensors to measure the entire vacuum range from UHV (10-9 Torr) to above the atmosphere (103 or 1000 Torr). This vacuum gauge has a variety of features, including RS-485 communication, two programmable setpoints, and a selectable analog vacuum pressure gauge output.
The bright LED digital display of the vacuum gauge is easy to read even from a distance. The options of this wide-range vacuum gauge can be adjusted via four touch controls on the front of the unit or via RS-485 communication. This vacuum gauge offers an extensive range of connections, easy operation, and easy sensor change, making it an excellent choice for a wide range of vacuum applications.
Operating range
1*10-9 Torr to 1000 Torr
Communication
RS-485
Analog Output
0 V to 10 V linear/logarithmic
Programmable setpoints
3 Relays
Supply voltage
22 V DC to 26 V DC
Maximum performance
8 W
Calibration medium
Dry air or nitrogen
Overpressure
150 PSI
Resolution of the digital output
2 significant digits with exponent
Analog output resolution
16 bits
Operating temperature
0 °C to 50 °C
Storage
-40 °C to 85 °C
Bake-out temperature
110 °C (electronics removed)
Reaction time
≤500 ms
Display Readable Display
4 m (13 ft)
Accuracy
10-9 Torr to 10-1 Torr
±0.2 of the decade
10-1 Torr to 200 Torr
±15% of reading
200 Torr to 1000 Torr
±5% of reading
Analog Output
±10 mV
PIN
Description
1
Analog Out (Ground)
2
RS-485 A (-)
3
SP3 Out
4
SP2 Out
5
SP1 Out
6
Zero adjust input (signal in)
7
EXT VCc (signal in)
8
Supply Voltage (+)
9
Analog Out (0 to 10) V
10
RS-485 B (+)
11
SP2, SP3 COM
12
SP1 COM
13
HV inhibit signal (signal in)
14
Atm adjust input (signal in)
15
Supply Voltage (-)
<STX>0R1<CR>
Read units
<STX>0R2<CR>
Read SP1L and SP1H
<STX>0R5<CR>
Read analog output set parameter
<STX>0W1XXXX<CR>
Set pressure unit
<STX>0W2XXXXXXXX<CR>
Set SP1L and SP1H
<STX>0C1<CR>
Crystal gauge atm adjustment
<STX>0S1<CR>
Read pressure data
<STX>0S2<CR>
Read error status
Assume address 0
Full list of commands is available in the instruction manual.
The CC-10 has three set points. All three are optically coupled solid state relays that are rated for 0.12 A at 50 V DC. Each set point has hysteresis characteristics and consists of two set values: upper threshold and lower threshold. LEDs on the front panel display the status of each set point. The high and low threshold values of each of the three set points can be set independently.
Molybdenum
Glass
Copper
Stainless Steel
Chromium
Solder
Fluorocarbon Elastomer
Quartz
Enclosure Material
Metallized plastic
Electrical connections
D-Sub 15-pin male
Weight with sensor
698 g (1.54 lbs) (NW25 Flange)
Weight without sensor
355 g (0.78 lbs)
CE certified to EN 61010-1, EN 61236-1
RoHS compliant
IP40
ISO 17025 accredited calibration optional
NIST-traceable calibration optional
Very Compact Wide Range Vacuum Pressure Gauge
Easy-to-read digital LED vacuum gauge display
Vacuum gauge with excellent resolution and repeatability
Superior instrument-to-device performance between vacuum gauges
ISO 17025 accredited and NIST traceable calibrations optional